3. Diagnostics and Troubleshooting
Cycle Fault Effect Problem Possible Causes
Touchscreen is blank/
white
Check power source
USB storage device
does not contain
the last print out
Re-insert the USB storage
device and wait for the
data to copy over again. If
problem persists, back up all
the information on the USB
device and reformat it.
NOTE: the web portal allows
access to all of the unit’s 
cycle information.
Unit is not sending
emails
Check email settings by
using the TEST button on
the unit’s web portal. 
From the SETUP web page,
select the TOOLS tab.
Click on TEST to check
your router, unit, and Internet
connections. If all settings
appear to be OK. Go to
the unit’s touchscreen 
and renew
the IP address by following
these steps:
1. Scroll through the setup
menu to NETWORK
SETUP and select.
2. Select RENEW IP.
Not receiving emails
from the unit
Check user’s spam lter. 
Be certain the unit has been
identified as an accepted
email source.