6.1 Replacing the HIP Cleaning Solution
The Hydrim L110w uses a 3.8L / 1 gallon container of HIP Cleaning Solution
(Part number CS-HIPL). Use only the recommended cleaning solution and read the
MSDS before inserting into the unit.
To order additional HIP Cleaning Solution, contact your local dealer. When the
message “no detergent” appears on the screen, the cleaning solution container is empty
and is in need of changing.To avoid spillage, replace the solution container only when
the message appears and be sure to wear gloves.
NOTE: Each cleaning solution container will last approximately 30 cycles,
depending on which cycles are used..
To replace the cleaning solution container, follow these steps:
1. Turn the power switch OFF.
2. Open the cleaning solution door.
3. Disconnect the cleaning solution connector. Remove any dried chemical by
immersing the connector in a cup of water.
4. Remove the empty cleaning solution container and discard or recycle.
5. Connect new container and insert into the cleaning solution compartment.
Follow the instructions printed on each cleaning solution container.
6. Close the door.
7. Turn the power switch ON
18
6. Care and Maintenance