5. Instructions for Use
5.2 Cycle Description Chart
10
* This is not a wash cycle. Always run a wash cycle following the rinse & hold cycle.
** Cycle times depend on the temperature and pressure of incoming water.
*** Rinse times are adjustable up to 10 minutes by an authorized technician.
Cycle Prewash Wash Rinse Dry
Total Time
**
w/o Drying
Water
Consumption
P0 – Machine
Cleaning Cycle
No initial draining.
<30ºC/86°F
(cold)
3 minutes
N/A
<30ºC (cold)
2 minutes
N/A
7
minutes
16 L
4.23 Gal
P1 – Rinse and
Hold Cycle*
Use to prevent soil from
drying on instruments
when they will not be
washed within one hour.
<30ºC/86°F
(cold)
3 minutes
N/A
60ºC/140°F
1 minute
N/A
17
minutes
16 L
4.23 Gal
P2 – Regular Cycle
(no disinfection) – Use for
moderately soiled loose
instruments.
N/A
50ºC/122°F
5-15 minutes
(default 5 minutes)
60ºC/140°F***
1-10 minutes
(default 1 minute)
1-20 minutes
(default 10 minutes)
25
minutes
24L
6.34 Gal
P3 – Heavy Duty
Cycle
Use for heavily soiled
instruments and cassettes.
<30ºC/86°F
(cold)
3 minutes
50ºC/122°F
5-15 minutes
(default 9 minutes)
60ºC/140°F
1 minute
1-20 minutes
(default 10 minutes)
34
minutes
32L
8.45 Gal
Page view 11