4. Installation Instructions
Testing water quality
The HYDRIM L110w G4 is equipped with
a built-in water softening system that must
be adjusted according to the local water
hardness. To read local water hardness,
proceed as follows:
1. The water test kit included with your
HYDRIM contains three water
hardness test strips in bags. Take a
water sample from the location where
the machine will be installed.
2. Open one of the bags, remove the
test strip and dip it into the water.
3. Compare the color of the strip with
the chart on the back of the bag
Determine the water hardness according
to the chart on the water test kit envelope.
4. Power the unit on and select the
Settings key from the main menu.
5. Go to the Setup Menu and select
“Set Regeneration”.
6. Using the up and down arrows, set the
water softener regeneration level
according to the water hardness table
in this section. If your water hardness
falls between two settings, select the
higher setting.
7. Unscrew the water softener container
lid from the bottom right of the chamber
and pour at least 1.0 litres (0.26 gal) of water
into the water softener container. Then fill the
salt container to the top (maximum 1 kg/
2.2 lbs.) with salt. Close by screwing the lid
tightly back into place and run a complete
cycle with no instruments. An improper seal can lead to corrosion.
*Please note: The water test strip is only accurate up to 250 ppm. If the reading on the test strip exceeds
250 ppm and/or if the location in which the HYDRIM is installed has known water quality problems, having
a more detailed and accurate water test done by a test lab is strongly recommended.
Water Hardness Conversion and salt regeneration levels
°dH
US
GPG
PPM
(mg CaCO3 / Litre)
Regen.
1 1.0 18
2 2.1 36
3 3.1 54
Very Soft
4 4.2 71
5 5.2 89
5.6 5.8 100
1
6 6.3 107
6.2 6.4 110
7 7.3 125
8 8.3 143
2
8.4 8.8 150
9 9.4 161
10 10.4 178
10.1 10.5 180
11 11.5 196
3
11.2 11.7 200
11.8 12.3 210
Typically
No Treatment
Necessary
(values from 18-143)
External Treatment
Required
(values >535)
Additional
Water
Treatment
Required
May Require External Treatment
(values from 150-535)
12 12.5 214
13 13.6 232
4
14 14.6 250*
15 15.6 268
16 16.7 286
5
16.8 17.5 300
17 17.7 303
18 18.8 321
19 19.8 339
6
19.6 20.5 350
20 20.9 357
20.2 21.0 360
21 21.9 375
22 22.9 393
22.4 23.4 400
23 24.0 411
24 25.0 428
25 26.1 446
7
25.2 26.3 450
26 27.1 464
27 28.2 482
28 29.2 500
28.6 29.8 510
29 30.2 518
30 31. 3 535
30.3
31.6
540
8
8