• PCB Rev. 7.0 Replacement InstructionsPCB Rev 7.0 Replacement Instructions 96-109424 Rev 3.0 2009 SciCan LtdSTATIM 2000/S/E & 5000/S
8 3. Statim Data Logger Quick Setup Guide For units with Rev. 7 Printed Circuit Board Only Note: Please do not connect Statim Data Logger. Follow i
9 Step 3: Setup Unit ID. a) Power up unit while holding the Stop button pressed. The user menu is going to be displayed. b) By using Unwrapped an
10 4. PCB Rev. 7.0 Assembly Instructions All the connections remain the same as per previous PCB revisions with the exception of the thermocouple co
5.PCB 7 Replacement Kits ST5000/S 11 1. 01-109370S PCB SEVEN 100V/110V ST5000 NA, Japan • PCB Seven + Software 2. 01-109746S PCB 7 Repl. Kit
5.PCB 7 Replacement Kits ST5000/S 12 16. 01-109842S PCB 7 Repl. Kit ST5000 230V, No Pr. • PCB Seven + Software + Serial Cable • Instructions
6. PCB 7 Replacement Kits ST2000/E/S 1. 01-109365S PCB SEVEN Alex 110V 2000 B: y PCB Seven + Software 2. 01-109690S PCB 7 Repl. Kit, ST2000 110V, B
14 7. Statim Autoclave Conductivity Circuit Calibration Instructions Statim Autoclave equipped with PCB rev 7.x requires Conductivity Circuit Calibr
158. Statim Rev. 7 PCB User and Service Menus 1. User setup menu...
16 1. User setup menu Note: To enter User Setup Menu power up Statim while Stop key is pressed To initially setup your Statim >Time/Date Set
171.2 Language Setup Display information in your desired language Display N.A ENGLISH Available Languages N.A ENGLISH JAPANESE U.K.ENGLISH SVEN
Manufactured by:SciCan Ltd.1440 Don Mills Road,Toronto ON M3B 3P9CANADAPhone: (416) 445-1600Fax: (416) 445-2727Toll free: 1-800-667-7733EU
181.4 Water Quality Display detected water quality Water Quality CD=xx.xuS/NNN/y.yppm Screen Representation xx.x Water conductivity in uS (micro-
191.7 End of Line CR/LF Configure the printout layout >End Of Line CR/LF CR/LF - CR This only needs to be set if a serial printer is attached
201.8 Serial Port Bit Rate Choose bit rate for device connected to the serial port >Serial Port Bitrate 9600 19200 57600 115200 300 1200 2400
212. Service Menu Note: To enter Service Menu power up Statim while holding Unwrapped and Wrapped keys pressed. After that enter password to acce
22 2.1. Calibration Calibrate thermocouple offsets 2.1.1. Non S units Chamber Calibration Screen VVV.V VO CCC.C CO
23Rubber and Plastics Start the Burn-in cycle Stop Exit to normal mode of operation Keypad functions when calibration cycle is running: Unwrapped
24To start the Chamber thermocouple Calibration cycle, first press and release the Unwrapped key followed by pressing the Start key. Wrapped Sends
25cycles and cycle parameters are displayed one by one on the LCD display. The available sets of cycle parameters are listed in the chart below. Av
26 Keypad Unwrapped Increment current field Wrapped Decrement current field Rubber and Plastics Move to next field Stop Exit Conductivity setup
27 2.8. Display last CF# Show the last Cycle Fault that occurred >Display last CF# ## (######) Screen Representation ## Value of last recor
11. Statim Non-S Calibration - Software revision 6.00 and higher Instructions for the Calibration of Statim Autoclave thermocouples Warning Î Inc
28 2.11. Press. Atm. Offset View the offset of the pressure sensor for atmospheric conditions >Press. Atm. Offset ## Screen Representation ##
292.14. Validation Offset View the offset of the thermocouple >Validation Offset ## Screen Representation ## Offset Value Keypad Rubber and
30 2.17. Repeater mode Enable or disable unit to run cycles continuously >Repeater mode On Off Keypad Unwrapped Select next option. Second li
31 2.20. Backup NVRAM Saves a copy of the unit’s current settings >Backup NVRAM No Yes Keypad Unwrapped Select next option. Second line shows
9. Statim 2000 Trouble-shooting Cycle Faults – Software Revision R1xx, R2xx, R4xx, R5xx and R6xx 32 Cycle Fault Number Description of Fault Proba
9. Statim 2000 Trouble-shooting Cycle Faults – Software Revision R1xx, R2xx, R4xx, R5xx and R6xx 33 Cycle Fault #6 Software Revisions R2xx or R4xx
9. Statim 2000 Trouble-shooting Cycle Faults – Software Revision R1xx, R2xx, R4xx, R5xx and R6xx 34 Cycle Fault #9 Not used Cycle Fault #10 The ca
9. Statim 2000 Trouble-shooting Cycle Faults – Software Revision R1xx, R2xx, R4xx, R5xx and R6xx 35 Cycle Fault #19 For software rev. R1xx, R2xx and
9. Statim 2000 Trouble-shooting Cycle Faults – Software Revision R1xx, R2xx, R4xx, R5xx and R6xx 36 Cycle Fault #26 Software Revision R5xx and R6xx
9. Statim 2000 Trouble-shooting Cycle Faults – Software Revision R1xx, R2xx, R4xx, R5xx and R6xx 37 Cycle Fault #98 Software Revisions R4xx and up (
2 5. Confirm that the display appears similar to the example below: The value in the lower right-hand corner (DDD) is the differenc
9. Statim 2000 Trouble-shooting Printer Messages 38 Cycle Fault Number Description of Fault Probable Cause of Fault Message "PRINTER FAULT&
9. Statim 2000 Trouble-shooting (Diagnosing Conductivity Sensor Problems for PCB Rev 2.x, 5.x and 6.x only) 39 STATIM / STATIM 1000 / 2000 UNITS CON
9. Statim 2000 Trouble-shooting Trouble-shooting, Original Statim Error Codes Unit Displays Select A Program 40 There are different types of Control
9. Statim 2000 Trouble-shooting Trouble-shooting, Original Statim Error Codes Unit Displays Select A Program 41 #5 between 2 minutes 44 seconds and
10. Statim 5000 Trouble-shooting – Cycle Faults 42 Cycle Fault Number Description of Fault Suggested steps for Correction of fault Cycle Fault #1
10. Statim 5000 Trouble-shooting – Cycle Faults 43 Cycle Fault #6 Software Revision R5xx and R6xx The software has detected the filtered Validation
10. Statim 5000 Trouble-shooting – Cycle Faults 44 Cycle Fault #13 N/A Cycle Fault #14 Software Revisions R1xx or R2xx only The steam generator te
10. Statim 5000 Trouble-shooting – Cycle Faults 45 Cycle Fault #25 The software has failed to detect a need to pump water within 90 seconds of the s
10. Statim 5000 Trouble-shooting – Cycle Faults 46 Cycle Fault #90 Software Revisions R6xx only Corrupted or not initialized chamber calibration val
10. Statim 5000 Trouble-shooting – Printer Messages 47 Printer Fault Description of Fault Suggested steps for Correction of fault Message: PRINTE
3 Validation Thermocouple Calibration: 11. New! Turn the machine off and back on, while keeping Unwrapped and Wrapped button pressed to go back
10. Statim 5000 Trouble-shooting – Printer Messages 48 is correctly connected if the time and date are shown on the LCD. Make sure that the printer i
11. Statim 2000S and 5000S Troubleshooting Cycle Faults – Software Revision R4xx, R5xx and R6xx 49 Cycle Fault Number Description of Fault Probab
11. Statim 2000S and 5000S Troubleshooting Cycle Faults – Software Revision R4xx, R5xx and R6xx 50 Cycle Fault #11 The cassette temperature failed
11. Statim 2000S and 5000S Troubleshooting Cycle Faults – Software Revision R4xx, R5xx and R6xx 51 Cycle Fault #28 The cassette pressure rose above
11. Statim 2000S and 5000S Troubleshooting Cycle Faults – Software Revision R4xx, R5xx and R6xx 52 Cycle Fault #55 For the Rubber and Plastics cycl
11. Statim 2000S and 5000S Troubleshooting Cycle Faults – Software Revision R4xx, R5xx and R6xx 53 Cycle Fault #70 The time maintained by the inter
11. Statim 2000 Trouble-shooting Printer Messages 54 Cycle Fault Number Description of Fault Probable Cause of Fault Message "PRINTER FAULT
55 12. Statim Data Logger Troubleshooting This troubleshooting addresses problems that may be encountered when operating the Data Logger. In th
56 T5. If MSD has: Corrupt or unreadable files and/or directories Refer to the STATIM screen to confirm successful sterilization. The M
57 13. Printer Interface Board rev 4.x Installation Instruction When replacing a non-Rev 7x PCB of a Statim 5000(S) with printer with a rev 7.x P
42. Statim S-Class Calibration - Software Revision 6.00 & Higher Instructions for the Calibration of Statim S-Class Autoclave Warning Î Inc
58 BOILER TC VALIDATION TC CHAMBER TC VALVE PUMP BOILER MAINS KEYPAD LCD LCD CONTRAST FLOAT CONDUCTIVITY PROBE CASSETTE SWITCH SERIAL COMMUNICA
54. New! Power up unit while keeping Unwrapped and Wrapped button pressed to enter Statim Service Mode. This Service Mode is password protected,
6CHAMBER CALIBRATION: (Temperature and Pressure) 6. To start a chamber calibration cycle press and release the Unwrapped cycle button and then
711.For software revisions higher than 4.15 (R415) when chamber pressure is higher than 115kPa the PO field will switch to indicate the POS – Pre